Rev Esp Quimioter 2016, 29(5):265-268

Correlation between MALDI-TOF Vitek-MSTM system and conventional identification methods of gastrointestinal infection causing bacteria                     

CARLOS RUIZ DE ALEGRÍA-PUIG, AMAIA AGUIRRE-QUIÑONERO, JESÚS AGÜERO-BALBÍN, Mª PIA ROIZ-MESONES, LUIS MARTÍNEZ-MARTÍNEZ          

Introduction. Rapid identification of pathogens is essential for the diagnosis of gastrointestinal infections. Matrix-assisted laser desorption/ionization time of flight (MALDI-TOF) mass spectrometry has shown to be effective and fast for the identification of microorganisms. The objective of this study was to evaluate the correlation between Vitek-MSTM and conventional methods for bacterial identification causing gastrointestinal infection.
Material and methods. A total of 329 gastrointestinal pathogens were identified using Vitek-MSTM (v2 SARAMIS MS -ID, bioMérieux, Marcy-I´Étoile, France) and routine diagnostic methods simultaneously. In cases of discrepancy 16SrRNA gene sequencing was performed.
Results. The correlation between Vitek-MSTM and diagnostic methods was 100% except for Yersinia enterocolitica (94.1%), Helicobacter pylori (10%) and Aeromonas veronii (0 %).
Conclusions. Vitek-MSTM is a quick and useful method for identification of enterophatogenic bacteria. It is necessary to improve the performance of the system for the identification of H. pylori and A. veronii.

Rev Esp Quimioter 2016; 29(5):265-268 [pdf]